### OBSDATE ### 20110413 ### LOGDATE ### 110413 ### OBSERVERS ### Barnaby Norris, Ben Warrington ### WEATHER ### 0 wind, ~55% RH. Clear. ### LOCATION ### SUSI ### PROGRAM ### Beta Car etc ### SIDEROSTATS ### n3 s3 ### TARGETS ### Beta Car (several) + calibrator HR 4199 Beta car + attempted calibrator HR 4037 - could not get fringes Beta car - save failed as data disk full ### PROBLEMS ### Siderostat controller n3 failed with `RTP read failed after 2 tries' Reset controller... Unable to see star in tip/tilt on PAVO, problem traced to shutter problem. The two (central pier vs. pavo pier) boards controlling the shutters seem somehow reversed, so incorrect shutters were moving. BW implemented temporary fix, will look at closely tomorrow. NOTE: Documentation does not describe how to save data! It assume autosave is on by default, which it is not. Needs to describe how to turn on autosave, etc. Warn the user that after shutters, pavo will stop - to the noob it's not clear when the shutter sequence is done. In a few cases, even when acq has acquired the star and is seen by pavo, taskmaster still does not move on to a fringe search. ERROR - taskmaster fails when trying to move pellicle, with message `Error in command acq'. and message `Pellicle move timed out'. run_acq;abjack has several errors - see log. Can still do it manually using buttons on acqlabjack gui. Cloudy! 75% RH now. Closing roofs. Roof move failed, due to UPS overloading and turning off. Ben reset it.